Bruel & Kjaer 4226 Specifications
B&K 4226 : Acoustic Calibrator - Class 1 - multi-function | n/a |
---|
Not available for hire, details retained for information purposes only.
Complies with IEC 942 Sound Calibrators, Class 1
at Reference Sound Pressure Level
at Reference Ambient Conditions
94 dB SPL ± 0.2 dB re 20 µPa, at 1 kHz
Nominal Sound Pressure Levels :
94 dB SPL, 104 dB SPL and 114 dB SPL10 dB and 20 dB Level Step Accuracy :
± 0.1 dB for f < 8 kHz; ± 0.2 dB for f > 8 kHzNominal Frequencies :
from 31.5 Hz to 16 kHz in octave steps, plus 12.5 kHzFrequency Accuracy :
± 1 % re the BS ISO 266 stated exact frequenciesFrequency Stability :
better than ± 30 ppmReference Ambient Conditions :
Ambient Temperature : 20 °C
Ambient Pressure : 1013 hPa
Influence of Ambient Conditions :
at Reference Sound Pressure LevelAmbient Temperature : +0.002 dB/°C in the range -10 °C to +50 °C, max. temp. uncertainty ± 5 °C
Ambient Pressure : +0.00055 dB/hPa in the range 650 hPa to 1080 hPa, max. pressure uncertainty ± 30 hPa to satisfy class 1 requirements
Magnetic Field Sensitivity : no observable effect < 0.01 dB at 100 A/m
Frequency Response :
Pressure Field : Linear 94 dB, 104 dB and 114 dB SPL and inverse A-weighting, 94 dB SPL at 1 kHz
Equivalent 0 ° Free and Random Field : Linear 94 dB SPL and inverse A-weighting, 94 dB SPL at 1 kHz
Level Accuracy at Fixed Frequencies :
Lin and inverse A-weightingPressure Field Response - re 1 kHz
31.5 Hz - 125 Hz | ± 0.15 dB |
---|---|
250 Hz - 500 Hz | ± 0.10 dB |
2 kHz - 4 kHz | ± 0.15 dB |
8 kHz | ± 0.25 dB |
12.5 kHz - 16 kHz | ± 0.50 dB |
Equivalent 0 ° Free and Random Field Response : re 1 kHz
31.5 Hz - 125 Hz | ± 0.20 dB |
---|---|
250 Hz - 500 Hz | ± 0.10 dB |
2 kHz | ± 0.20 dB |
4 kHz | ± 0.30 dB |
8 kHz | ± 0.50 dB |
12.5 kHz | ± 1.00 dB |
16 kHz | ± 1.50 dB |
Total Harmonic Distortion :
< 1 %Additional Functions :
Time Weighting Tests :
Frequency : 2 kHz pre-selected
Test Level : Typically 94 dB SPL, adjustable ± 12 dB approximately
Signal : Continuous reference and tone burst
Level between Bursts : Ref. level -20 dB
Burst Duration : 500 ms Slow and 200 ms Fast
Crest Factor Test :
Frequency : 2 kHz pre-selected
Test Level : typically 94 dB SPL, adjustable ± 12 dB approximately
Signal : continuous reference and tone burst of 40 Hz repetition frequency. Crest Factor 3 at pre-selected and higher frequencies
Note :
All burst signals consist of an integral number of sine waves with amplitude shifting at zero crossingDose Meter Calibration :
Calibration Period : 30 s : Obtained doses at 1 kHz:
SPL | IEC | ANSI |
---|---|---|
94 dB SPL | 0.26 % ± 0.01 % | 0.18 % ± 0.01 % |
104 dB SPL | 2.64 % ± 0.02 % | 0.73 % ± 0.03 % |
114 dB SPL | 26.7 % ± 0.02 % | 2.90 % ± 0.12 % |
AC Output :
Output Voltage : 12.5 mV ± 0.5 mV at settings 94 dB, 1 kHz and pressure mode.
Voltage changes according to the selected level, type of microphone , sound field and function. Signal for testing time weighting and crest factor capabilities as above, except that the crest factor is 10
Note:
The acoustic output signal is switched off when a plug is inserted in the AC OUT socketOutput Impedance :
600 Ω short circuit proofFrequency Response : ± 0.2 dB from 31.5 Hz to 16 kHz in pressure mode
Note : The acoustic output is switched off when a plug is inserted in the AC OUT socket
External Generator Input :
Input Voltage : 0.5 V to 1.5 V
Input Impedance : 47 kΩ
Frequency Response in the Coupler :
31.5 Hz - < 8 kHz | ± 0.15 dB |
8 kHz - < 12.5 kHz | ± 0.25 dB |
12.5 kHz - 16 kHz | ± 0.50 dB |
Note:
These tolerances are relative to the inverse pressure response of the reference microphone. Only Pressure and Linear response are possible with external generator.Stabilization Time:
The LED corresponding to the selected frequency lights continuously when the SPL is stabilized, typically 2 to 10 secondsPower Supply :
Batteries : 4 x Lr6 - 1.5 V alkaline
Battery Life : 10 hours approximately at 20 °C with alkaline batteries
Battery Check : The 'Batt Low' LED is lit when reaching the last 10 % of the battery life
Effective Coupler Volume :
Approximately 400 cm3 at 31.5 Hz, decreasing to approximately 30 cm3 at 1 kHz, obtained by feedbackEnvironmental :
Ambient Temperature : -10 °C to +50 °C
Ambient Pressure : 650 hPa to 1080 hPa
Ambient Humidity : 5 % to 90 %RH
Dimensions and Weight :
including mahogany caseLength : 265 mm
Width : 125 mm
Height : 62 mm
Weight : 1.5 kg